Module 6: Particles and medical physicsUsing ultrasound (6.5.3)

Using ultrasound (6.5.3)

Ultrasound, the piezoelectric effect, transducers, acoustic impedance, A-scans and B-scans, and ultrasound imaging in A-level Physics.
9 min

Sound waves are a type of mechanical wave. When a sound wave propagates through a medium, the particles oscillate in a direction parallel to the wave’s propagation and energy transfer. Hence, sound waves are categorised as longitudinal waves.

An illustration showing the propagation of wave from a speaker to an ear. The image includes labeled sections for Compression, Rarefaction, and Oscillation of particle, with arrows indicating the direction of wave propagation.

A longitudinal wave is a series of compressions, where the particles are close together, and rarefactions, where the particles are spread out.

Add to favourites

Sound waves can have frequencies ranging from a few hertz (Hz) to multiple gigahertz (GHz).

Humans can only hear sound between and this range is called audible sound.

Sound waves that lie below the audible range (i.e. below ) are called infrasound, and the ones that lie above the audible range (i.e. above ) are called ultrasound.

A diagram illustrating sound waves categorized by frequency. The left section labeled 'Infrasound' shows waves at 20 Hz, the middle section labeled 'Audible sound' shows waves at 20 kHz, and the right section labeled 'Ultrasound' shows waves. The horizontal axis is labeled 'Frequency'.
Add to favourites

The generation and detection of ultrasound waves, mainly used by medical devices, depend on the piezoelectric effect.

The piezoelectric effect is the ability of certain piezoelectric crystals, such as quartz and certain ceramics, to generate an electric potential difference in response to mechanical stress (compression or extension).

The effect is reversible. So, a piezoelectric crystal will deform when a potential difference is applied.

Stress is applied ↔ Potential difference is observed. Compression. Piezoelectrical crystal. Compression. Potential difference is applied ↔ Stress is observed. Compression. Piezoelectrical crystal.
  • When stress is applied to a piezoelectric crystal, mechanical energy is converted into electrical energy.
  • When a potential difference is applied to a piezoelectric crystal, electrical energy is converted into mechanical energy.
Add to favourites

The diagram below shows the molecules of a piezoelectric crystal.

An illustration showing a piezoelectric crystal on the left, with arrows indicating compressing force in the middle, and stretching force on the right.
  • Under no stress, the molecules are randomly oriented and no charge is observed at either end of the crystal.
  • When the crystal is compressed, the molecules orient themselves so that the crystal becomes electrically polarised.
  • However, when the crystal is stretched, the polarisation is reversed.

For example, in the diagram above, compressing the piezoelectric crystal may induce a positive charge at the upper end of the crystal and a negative charge at the lower end, while stretching the crystal has the opposite result.

Add to favourites

When a power source is used to apply an alternating potential difference to the ends of the piezoelectric crystal, it compresses and expands so that the molecules of the crystal align with the external electric field, as shown below.

An illustration showing two states of a material: on the left, 'Compression' with arrows indicating force direction and an electric field vector E; on the right, 'Expansion' with similar arrows and the electric field vector E.
  1. Applying an alternating voltage at frequency to the piezoelectric crystal would compress and expand, or vibrate, the crystal at that frequency.
  2. As the crystal vibrates, it disturbs the surrounding air particles, thus producing sound waves of frequency

By applying an alternating potential difference with a frequency greater than ultrasound waves are produced.

Add to favourites

A device that converts energy back and forth between mechanical energy and electrical energy by utilising the piezoelectric effect is called a piezoelectric transducer.

A piezoelectric transducer that is used in ultrasound imaging is shown below.

An illustration of a medical ultrasound transducer showing labeled parts including Coaxial cable, Insulating case, Backing layer, Piezoelectric crystals, and Acoustic lens.

A transducer consists of a thin slice of piezoelectric crystal protected by an insulating case. An acoustic lens at the front is used to converge the ultrasound beam.

The crystal is backed by a damping layer, usually epoxy resin, to quickly stop its vibrations after emitting the ultrasound pulse. This reduces interference between the emitted and reflected pulses.

To maximise the efficiency of a transducer, the frequency of the waves must match the resonant frequency of the crystal. The thickness of the crystal should be about half the wavelength of the emitted ultrasound.

  • To generate ultrasound, the transducer converts an alternating potential difference into sound waves.
  • To detect ultrasound, the transducer converts the sound waves into an alternating potential difference.
Add to favourites

The simplest type of ultrasound imaging is called an A-scan, or amplitude scan:

  • They utilise a single transducer to emit a signal and subsequently receive the reflected signal.
  • A-scans are one-dimensional, determining the depth of internal structures.

The depth to the subject of the scan from the skin is calculated by measuring the time delay between the signal generation and reception.

An illustration showing an eye with a probe, labeled 'Voltage' on the left side and 'Time' on the bottom, with a graph depicting voltage changes over time.

In A-scans, ultrasound waves are typically reflected at the boundary between two media.

Examples of A-scan usage include:

  • measuring the length of the eye,
  • identifying the presence of tumours.

A-scans provide depth measurements only and do not produce images.

Add to favourites

The depth or distance to a target in the body can be calculated using an A-scan.

For a time delay between the generation and reception of the ultrasound, and by knowing the speed of ultrasound in the tissue, the depth of the target can be calculated from:

The factor of is due to how the ultrasound pulse must travel from the transducer to the boundary of the target and then back again, so the measured time corresponds to twice the one-way distance.

An illustration showing an eye with a probe, labeled 'Voltage' on the vertical axis and 'Time' on the horizontal axis. The graph depicts voltage changes over time with peaks and troughs, and the symbol 'Δt' indicating a time interval.
Add to favourites

A B-scan, also known as a brightness scan, is an ultrasound imaging technique that generates two-dimensional or three-dimensional images of internal body structures. B-scans use a moving sensor or multiple sensors to capture ultrasound waves.

The transducer is moved across the patient’s skin, or multiple transducers are employed to acquire a series of A-scans, which are then combined to create detailed images.

Add to favourites

A B-scan image shows the reflected ultrasound intensity at each point on the sensor; for example, high-intensity ultrasound appears as white, while no reflected signal appears black.

An ultrasound image showing a fetus in a womb, with a heart shape visible.
  • B-scans are primarily used for diagnostic purposes, such as visualising bones, muscles, and organs or monitoring a foetus’s development.
  • To enhance image clarity, pulsed ultrasound waves are employed to ensure that reflected waves are received without interference.
  • Shorter wavelengths are used to provide greater detail, allowing sound waves to navigate around finer internal structures.
Add to favourites

The table below highlights the key differences between an A-scan and a B-scan.

A comparison table showing A-scan and B-scan ultrasound techniques. A-scan: A one-dimensional ultrasound scan that measures the distance to internal structures. Displays a single line of data, showing amplitude versus time. Used to measure the length of the eye's internal structures. Typically employs a stationary transducer that focuses on one point. B-scan: A two or three-dimensional ultrasound scan that creates images of internal structures. Produces a cross-sectional image. Used for visualising organs, tissues, and monitoring foetal development. Involves moving the transducer to capture data at different positions.
Add to favourites

Question walkthrough

Calculating Eye Length from an A-Scan

Use the pulse-echo time interval and speed of sound in tissue to calculate the length of an eye from an ultrasound A-scan.

Acoustic impedance is a measure of how much resistance an acoustic wave encounters as it travels through a medium.

The acoustic impedance of a substance is defined as the product of the density of the substance and the speed of the ultrasound in the substance:

The SI unit of acoustic impedance is The quantity is the speed of sound in the given substance, not the speed of light.

Add to favourites

The speed of sound in a material depends on both the elasticity and the density of the material.

  • Gases have low densities, and ultrasound travels more slowly in gases than in other states of matter. Therefore, their acoustic impedance is low.
  • Conversely, solids are generally denser, and ultrasound travels fastest in solids. As a result, the acoustic impedance of solids is high.
A table displaying various substances along with their phase, density (kg m-3), impedance (kg m2 s-1), and speed of sound (ms-1). The substances listed include Air (Gas, 1.23, 414, 343), Nitrogen (Gas, 1.25, 412, 334), Oxygen (Gas, 1.43, 484, 316), Ethanol (Liquid, 789, 1,200, 1,161), Oil (Liquid, 850, 1,300, 1,420), Water (Liquid, 1000, 1,480, 1,482), Aluminium (Solid, 2,700, 70,000, 5,100), Steel (Solid, 7,850, 43,000, 5,960), and Copper (Solid, 8,960, 42,000, 3,850).
Add to favourites

Question walkthrough

Comparing Acoustic Impedance of Air, Blood and Bone

Rank the speed of ultrasound and acoustic impedance across air, blood, and bone, then calculate bone's acoustic impedance.

When a collimated beam of ultrasound is incident on a boundary between two media, it is partially reflected, and the rest is transmitted into the next medium.

The ratio of the reflected beam intensity to the incident beam intensity depends on the acoustic impedances, and of Medium 1 and Medium 2.

A diagram illustrating the behavior of light at a boundary between two media. The top section labeled 'Medium 1' shows an 'Incident beam' directed downward, with a 'Reflected beam' directed upward. The bottom section labeled 'Medium 2' shows a 'Transmitted beam' directed downward, with a 'Boundary' separating the two media.
Add to favourites

When the incident ultrasound beam is perpendicular to a boundary between two media, the relationship between the reflected beam intensity incident beam intensity and acoustic impedances, and is given by:

The ratio is called the intensity reflection coefficient.

The SI unit of intensity is

  • The greater the difference between the acoustic impedances of the two media, the greater the reflected intensity, resulting in a smaller portion of the beam being transmitted.
  • If the acoustic impedances are the same, the reflected intensity is zero. This is equivalent to the ultrasound wave travelling through a single material with no boundary.
Add to favourites

The transmitted intensity is equal to the incident intensity minus the reflected intensity

The percentage of the initial intensity that is transmitted, added to the percentage of the initial intensity that is reflected, equals 100%:

Add to favourites

Question walkthrough

Reflection and Transmission at a Fat-Bone Boundary

Calculate the percentage of an ultrasound beam reflected and transmitted at a fat-bone tissue boundary.

During an ultrasound examination, a transducer is typically placed on the patient’s abdomen or chest. If air pockets are trapped between the transducer and the skin, a significant difference in acoustic impedance is created.

As a result, a large portion of the incident ultrasound beam is reflected back, with only a small amount transmitted into the body. This can make studying the internal structure of the human body difficult.

To counteract this effect, a special gel, known as a coupling gel, is placed between the skin and the transducer. An example of this is shown in the diagram below, during the study of a lesion.

An illustration showing the layers of skin, fat, muscle, and a lesion with a tumour. The transducer is positioned on the skin with coupling gel applied.
Add to favourites

The coupling gel used in ultrasound examinations has an acoustic impedance very close to the value of the skin.

  • It eliminates air gaps and enhances the transmission of ultrasound into the body.
  • When a coupling gel is used, we say we are applying impedance matching.
  • When a coupling gel is used, reflection at the boundary is minimal.

From the equation below, the reflected intensity is equal to zero when Therefore, the closer the values of the acoustic impedances of the two media are, the smaller the reflected intensity will be.

Add to favourites

Question walkthrough

Choosing Coupling Gel to Minimise Ultrasound Reflection

Compare acoustic impedance mismatches to identify which coupling gel minimises ultrasound reflection at the skin boundary.

When ultrasound reflects off a moving object, its frequency changes. This is known as the Doppler effect.

Doppler ultrasound is a non-invasive medical technique that utilises ultrasound reflections from iron-rich blood cells. This helps doctors evaluate blood flow through major arteries and veins, including those in the arms, legs, neck, and heart.

An illustration showing a transducer with a transmitting element and a receiving element positioned above tissue and gel. Below, there is a depiction of blood flow with red blood cells and an angle θ indicated.

Applications of Doppler ultrasound:

  • Reveal blood clots.
  • Identify the narrowing of vessel walls resulting from the accumulation of fatty deposits.
  • Evaluate blood flow to transplanted kidneys or livers.
Add to favourites

To perform a Doppler ultrasound:

  1. A transducer is gently placed on the skin above the blood vessel.
  2. The transducer transmits ultrasound pulses and captures the reflected signals.
  3. Reflections from moving blood cells show a change in frequency – the frequency increases when blood flows toward the transducer and decreases when it flows away.
  4. The frequency shift is directly proportional to the speed of the blood in the vessel.
  5. The transducer is connected to a computer that creates a colour-coded image, illustrating the direction and speed of blood flow on a display.
Add to favourites

In Doppler ultrasound, the change in frequency as a fraction of the initial frequency is equal to:

Where:

  • is the speed of blood,
  • is the speed of the ultrasound, and
  • is the angle between the ultrasound receiver and the line along which the blood is travelling.
An illustration showing a transducer with a transmitting element and a receiving element positioned above a blood vessel. The image includes labels for tissue, gel, and blood flow, with an angle θ indicated in relation to the blood flow.
Add to favourites

Question walkthrough

Doppler Ultrasound Blood Flow Speed

Calculate the speed of blood flow from the Doppler frequency shift of reflected ultrasound.