Section III - Reasoning in Biological and Physical SciencesScientific literacyPhysicsMedical physics

Medical physics

Apply physics principles directly to healthcare contexts, especially imaging and radiation. Cover X-rays/CT/MRI/ultrasound at a conceptual level, plus dose, attenuation, and effects.
26 min

X-rays can be artificially produced using X-ray tubes. Electrons are expelled from a heater (cathode) into a vacuum via thermionic emission:

  1. Thermal energy from the heater liberates electrons from their atomic nuclei.
  2. The electrons are then rapidly accelerated by a high voltage electric field through the vacuum towards a target metal.
  3. Finally, the electrons are rapidly decelerated by colliding with a target metal anode (e.g. tungsten).
  4. Colliding with the anode makes the electrons lose most of their kinetic energy. The lost kinetic energy is converted into X-rays.
A diagram of an X-ray tube showing various components labeled: Oil for heat conduction, Lead casing, Electrons, Glass envelope containing vacuum, Anode, Target, Window, X-rays, Filament, Focusing cup, and Cathode.

It is important to note that the internal vacuum is necessary to prevent electrons from colliding with air molecules. An external power supply creates a potential difference across the cathode and anode. A high-voltage supply is required to generate an electric field strong enough to produce X-rays.

Add to favourites

Inside an X-ray tube, a filament (cathode) is heated so electrons gain energy; if they gain enough energy to escape the metal surface, this is called thermionic emission. The emitted electrons cross the vacuum and are accelerated towards a metal target (the anode).

When the electrons strike the anode, only about 1% of their kinetic energy is converted into X-ray photons, with the rest transferred to the anode as thermal energy.

Add to favourites

X-rays are produced in every direction from the anode of an X-ray tube. However, in practice, the X-rays need to be guided in one direction towards a sample or object, such as a suspected broken bone. Hence, one side of the X-ray tube has a window made of a material thinner than the lead casing to allow unidirectional emission of X-rays.

Diagram of an X-ray tube showing components including Oil for heat conduction, Lead casing, Electrons, Glass envelope containing vacuum, Anode, Target, Window, X-rays, Beam collimator, Filament, Focusing cup, and Cathode.

The beam is directed into a series of straight, parallel metal tubes (called a collimator) that further confine the beam by absorbing any rays that are not parallel to the axis of the tubes.

Add to favourites

Inside an X-ray tube, a heated filament (cathode) emits electrons via thermionic emission. The filament is typically made of tungsten, which can withstand high temperatures. A high external voltage of around is applied between the cathode (negative) and the anode (positive), creating an electric field. This accelerates the emitted electrons towards the anode.

An illustration of an X-ray tube showing various components including a vacuum, tungsten target, filament, electrons, and X-ray emission. The diagram includes arrows indicating the direction of electron movement and electric field.

A potential difference translates to a kinetic energy gain of of the electrons. It is important to note the difference in units here. This is due to the equation:

Where:

  • is the energy transferred to the particle,
  • is the charge (an electron has a charge of one electron volt eV), and
  • is the potential difference across the X-ray tube.
Add to favourites

When a high-speed electron is incident on the tungsten anode in an X-ray tube, X-rays are produced by two different types of interactions:

Bremsstrahlung radiation: Electrons slow down or brake when they come near the nucleus of the tungsten atoms, they lose energy in the form of X-ray photons. This type of radiation produces a broad spectrum of X-rays.

It is a broad spectrum as the electrons can have a range of kinetic energies as they accelerate through the X-ray tube. This creates a range of kinetic energy losses when they collide with the anode. Bremsstrahlung radiation is sometimes referred to as braking radiation.

A graph showing the relative intensity of X-rays plotted against photon energy (keV). The curve peaks and then declines, with labels indicating 'Characteristic radiation' and 'Maximum photon energy'.

Characteristic radiation: Some electrons collide with the inner shell electrons of tungsten atoms, knocking them out of their orbit. When electrons from higher energy levels drop down to fill the vacancy, they release energy in the form of X-ray photons with specific, characteristic energies.

Add to favourites

X-rays ionise the matter they pass through, and will lose energy when interacting with atoms or molecules. As the X-ray passes through matter, the intensity (or energy) of the beam will decrease: this is known as X-ray attenuation.

There are four main X-ray attenuation mechanisms that cause X-ray beams to be attenuated and the X-rays to lose energy:

  1. Simple scattering
  2. Photoelectric effect
  3. Compton scattering
  4. Pair production

The mechanism that dominates is dependent on the energy of the incoming X-rays.

Add to favourites

Simple scattering is a type of X-ray attenuation mechanism and is common in the lowest energy X-rays in the range of

An incoming X-ray photon will be reflected by an electron in the material. The X-ray will be elastically scattered and change direction.

An illustration showing an atom with labels for Incoming X-ray, Scattered X-ray, Electron, and Nucleus. The atom is depicted with concentric circles representing electron orbits and a cluster of colored spheres representing the nucleus.

This is the simplest attenuation mechanism, and will not affect the energy or wavelength of the X-ray beam. It will not lose energy as the incoming X-ray has insufficient energy to ionise the atom.

Add to favourites

The photoelectric effect is a type of X-ray attenuation mechanism that affects X-rays with energy less than but greater than

These X-rays can match the ionisation energy of the atoms in the material. This results in the incoming X-ray being absorbed by an orbiting electron and the emission of a photoelectron (a liberated electron from within the atom).

X-ray is absorbed and releases electron. Incoming X-ray leads to a photoelectron. The atom is ionised, creating an electron vacancy. Outer electron relaxes down, which releases a photon.

After the photoelectron leaves the atom, another electron may transition down to fill the now vacant position: this electron movement also releases a scattered low-energy photon as the electron moves from a higher energy state to one of lower energy.

Add to favourites

The Compton effect is an X-ray attenuation mechanism which dominates for X-rays with energies between and

An incoming X-ray photon will inelastically scatter off an electron within a material: it is not absorbed by the atom as it has too much energy. Instead, inelastic scattering imparts some of the energy into the atom. The scattered X-ray photon will have decreased energy, and therefore an increased wavelength.

An illustration showing the interaction of X-rays with atoms. It includes an 'Incoming X-ray' represented by a wavy orange line, a 'Compton electron' indicated by a small yellow circle, and a 'Scattered X-ray' also depicted as a wavy orange line. The central part of the image shows a cluster of particles, with some colored pink and purple, representing atomic structure.

The electron involved (sometimes referred to as a Compton electron) will also be ejected from the atom, scattered in a different direction in order to conserve momentum.

Add to favourites

Pair production is an X-ray attenuation mechanism that can occur for the highest energy X-rays over

An incoming X-ray photon will interact with the electric field of an atom and spontaneously transfer all energy into producing an electron–positron pair that travels in opposite directions to conserve momentum. The positron will then quickly collide with a different electron and annihilate. Annihilation of a particle–antiparticle pair will produce photons.

The incoming X-ray photon must have a lot of energy to produce an electron–positron pair; these energies are not used in medical imaging, so pair production normally is not observed in this context.

An illustration showing an incoming X-ray interacting with an atom. The atom consists of a nucleus made up of pink and purple spheres, representing protons and neutrons, with yellow spheres surrounding it, representing electrons. The image indicates the emission of a positron and an electron as a result of the interaction.

It is useful to note that Compton scattering is the main cause of X-ray energy loss between roughly 30 keV and a few MeV. This is because its interaction probability decreases slowly. Other processes like photoelectric absorption (at lower energies) and pair production (above 1.02 MeV) also happen, but Compton scattering is simply the most likely to occur in that middle energy range. Remember that these “ranges” overlap; they aren’t strict cut-offs.

Add to favourites

X-rays are a form of ionising radiation: as they pass through matter they will lose energy due to attenuation. This causes a gradual decrease in the intensity of the X-ray, where intensity is defined as the power per unit of cross-sectional area and has units of

An illustration showing X-ray photons interacting with an object. The image includes labels for 'Photon energy', 'Attenuation', 'Penetration', and 'Thickness, density, atomic number (Z)'.

The penetrating power of the X-ray is dependent on the material it is passing through. An incoming X-ray will be attenuated based on:

  • the thickness of the material,
  • the density of the material, and
  • the atomic number of the material.

A very thick, dense, high atomic number material will attenuate the X-ray a lot more than a thin, less dense, low atomic number material.

Different materials will attenuate X-rays to a different extent; this can be used to distinguish different materials inside the body such as bone and soft tissue.

Add to favourites

The intensity of a collimated beam of X-rays decreases exponentially when attenuating. The intensity can be modelled as:

Where:

  • the initial intensity before the material is
  • is the intensity after the beam passes through a material thickness of (in metres), and
  • is the attenuation coefficient (units of ).

The attenuation coefficient (sometimes called the absorption coefficient) depends on the material the X-rays are passing through: it is a measure of how well a material can absorb X-rays. For instance, the attenuation coefficient is much higher for bone than soft tissue. The attenuation coefficient of materials is linked to the proton number,

Add to favourites

Applying the varied attenuation abilities of different materials is vital for medical imaging. For example, an X-ray image must discern between muscle, soft tissue, and bone to diagnose potential health issues in a patient.

Bone attenuates X-rays more than soft tissue. Therefore, X-rays propagating through bone are more likely to be absorbed, and the beam is more attenuated on the other side of the bone.

An X-ray image of a human hand showing the bones of the fingers and wrist.

Photographic film will turn from white to black if exposed to a high intensity of X-rays. Therefore, X-rays passing through bone will be attenuated and not turn the areas of the plate immediately behind the bone black. This contrast is evident in the image above. Modern digital detectors can also be used to enhance image processing, store and transfer images.

Add to favourites

Contrast media are easily seen on X-ray images; these materials have a high proton number , and hence a high attenuation coefficient. Common contrast materials:

  • barium
  • iodine (

Soft tissue has a composition of different atoms, which has an average proton number, , and has a low attenuation coefficient.

The correlation between the attenuation coefficient of a material and proton number is due to the photoelectric effect: this is the dominant absorption mechanism at X-ray frequencies of In general there is a proportionality of:

Contrast media are used to distinguish between different tissues in the body that look very similar in normal X-ray imaging.

Add to favourites

In medical X-ray imaging, barium and iodine are commonly used as contrast media.

Iodine is used as a contrast medium in liquids, for example, to observe blood flow. An organic iodine compound is injected into the bloodstream. In the X-ray image, low attenuated intensities indicate where the blood is flowing well, and areas with high attenuation (where the iodine is collecting) indicate poor blood flow.

Non contrast CT scan on the left and iodine contrast CT scan on the right, both showing abdominal images with arrows indicating specific areas of interest.

Barium sulphate is used as a contrast medium in the digestive system. A liquid ‘barium meal’ is swallowed by the patient. Similarly, areas where the barium builds up will be highly attenuated and can locate blockages.

Add to favourites

A computed axial tomography scan or CAT scan is a medical imaging technique that uses X-rays to analyse a patient’s body structure.

It comprises the following key elements:

  • Rotating X-ray source: an X-ray tube rotates around the patient. It produces a thin, fan-shaped X-ray beam that passes through the body, targeting specific slices or sections.
  • Ring of detectors: a ring of detectors encircling the patient capture the X-rays after they pass through the body. As the X-ray beam rotates, the detectors measure the amount of X-rays absorbed by different tissues, which varies based on the attenuation ability of different materials.
An illustration of a medical imaging device showing a person lying on a motorized table. The X-ray source is positioned above the person, and there are detectors indicated on the sides.

To generate images on a computer, software and displays are used.

  • Computer software: data from the detectors is sent to a computer. Using specialised software, the computer processes this data, performing complex mathematical calculations to generate detailed images of the scanned body part.
  • Display: processed data is converted into visual images that can be displayed on a monitor. These images show detailed cross-sections (slices) of the body, allowing healthcare professionals to make accurate diagnosis.
Add to favourites

There are advantages of using CAT (computed axial tomography) scans over traditional X-ray medical imaging:

  1. Cross-sectional imaging: CAT scans provide detailed cross-sectional slices of the body, offering a better view of internal structures. X-ray images are 2D, which can make it difficult to distinguish overlapping tissues.
  2. Better soft tissue contrast: CAT scans are superior at visualising soft tissues, such as the brain and blood vessels, which are hard to differentiate on X-ray images.
  3. Diagnostic accuracy: CAT scans can provide more extensive information for healthcare professionals to use for diagnosis, making their conclusions more accurate.
An illustration comparing a CT Scan of the brain on the left and an X-ray of the skull on the right. The text 'CT Scan' is above the brain image, and 'X-ray' is above the skull image.
Add to favourites

A radioisotope (or radionuclide) is an unstable isotope of an element that undergoes radioactive decay, emitting alpha, beta, or gamma radiation to achieve a more stable state.

An illustration of a radioisotope showing a cluster of spheres representing neutrons and protons. Arrows point to three types of radiation: Alpha radiation with a helium nucleus, Beta radiation with a high energy electron, and Gamma radiation with a high energy photon.

Radioisotopes used in medical imaging must have a short half-life to ensure high activity and be detectable within the body. This allows small amounts of the radioisotope to be used to create an image, ensuring patient exposure is kept to a minimum.

Add to favourites

Gamma-emitting isotopes are ideal for medical imaging due to their low ionisation and ability to penetrate the patient. Alpha and beta sources are not used in medical imaging because they are highly ionising, they can damage the cells in the body and lead to cancer.

Two common examples of radioisotopes used in medicine are:

  • Fluorine-18: Produced in hospitals using a particle accelerator and used in PET scans.
  • Technetium-99m: Produced by natural decay of molybdenum-90. Used to monitor the function of major organs, e.g. heart, lungs, brain, liver and kidneys.

The image below shows the molybdenum-90 decay chain where technetium-99m is produced:

90 42 Mo 67 h → 99 43 Tc + 0 -1 e + ν̅ e 6h 99 43 Tc + γ 210,000 y 99 44 Ru + 0 -1 e + ν̅ e

Molybdenum-90 beta-minus decays to form technetium-99m. The ‘m’ refers to metastable, meaning the nucleus remains in a high-energy state. Technetium-99m then beta-minus decays to technetium-99, which has a very long half-life of 210,000 years.

Add to favourites

To ensure the radioisotopes reach the desired location, they are attached to a targeting molecule using a linker (a chemical group that joins the radioisotope to the targeting molecule), which binds to the target protein on a cell membrane’s surface.

The combination of a radioisotope, linker, and targeting molecule is known as a radiopharmaceutical or a medical tracer. The diagram below shows the basic structure of a medical tracer:

An illustration showing a radiopharmaceutical targeting cancerous cells. The components labeled include: Radioactive compound, Linker, Targeted molecule, Target protein, and Cancerous cells.

It is useful to note that medical tracers can “label” a target in different ways, depending on the target molecule’s location. Some tracers bind to cell-surface receptors, whilst others are absorbed into cells by crossing the cell surface membrane. In both cases, the tracer accumulates more in the target area, making it easier to detect.

Add to favourites

Medical tracers can be injected into the bloodstream, swallowed, or inhaled, depending on the organ being targeted and the type of tracer used. Once inside the body, they travel to the intended location, where the radioisotope emits radiation detectable by imaging equipment such as a PET or gamma camera.

An example of a medical tracer is fluorodeoxyglucose (FDG), a compound made by combining the radioisotope fluorine-18 with a glucose-like molecule. FDG is used in PET scans to monitor metabolic activity, especially in the brain and cancerous tissues, as these areas absorb more glucose.

Add to favourites

A gamma camera is used to diagnose conditions by detecting gamma photons emitted from a patient’s body after they have been injected with a medical tracer. It monitors the function of various organs, such as the heart, lungs, liver, kidneys, and brain. It helps identify abnormalities such as reduced blood flow, blockages, or cancerous tumours by showing areas with abnormal tracer concentration.

For example, technetium-99m can be traced through the body using the gamma camera. Technetium-99m emits gamma photons, that can be detected as they exit the body, allowing the location of the radioisotope to be determined. Organs with lower concentrations of the medical tracer typically signify malfunction.

It is useful to note that technetium-99m can be combined with sodium and oxygen to create , a compound that targets brain cells and is used in brain imaging. Different medical tracers are designed to target different parts of the body.

Add to favourites

A gamma camera detects gamma photons emitted by medical tracers (usually containing Technetium-99m) injected into a patient. The diagram below shows the main components of a gamma camera:

An illustration showing a computer connected to electrical circuits, photomultiplier tubes, a light guide, a scintillator crystal, and a collimator. Below, there is a representation of a patient with an organ emitting gamma photons.

When a gamma photon passes through the collimator and strikes the scintillator crystal (often sodium iodide), it may produce a brief flash of visible light. These light photons are detected by photomultiplier tubes, which convert the faint light into an electrical signal and amplify it.

The signals are then processed by the electrical circuits and sent to a computer, which uses them to work out where in the scintillator each gamma photon was detected. By repeating this many times, the computer builds up an image showing the distribution (concentration) of the medical tracer in the body.

Add to favourites

The lead collimator in a gamma camera helps produce a sharp, high-resolution image by controlling the direction of incoming gamma rays. Without a collimator, gamma rays would enter from all directions, leading to a blurred, low-resolution image. The diagram below shows the basic structure of a collimator:

An illustration showing a Scintillator and a Collimator at the top, with red arrows indicating gamma photons emitted from an Organ emitting gamma photons below, and a label for Patient. The text 'Off-axis gammas are absorbed' is also included.

The collimator contains an array of small holes, allowing only gamma rays travelling along specific paths to reach the detector. It works by absorbing off-axis gamma rays, preventing them from contributing to the image:

  • The smaller the holes in the collimator, the higher the resolution, but the lower the sensitivity.
  • The larger the holes, the lower the resolution, but the greater the sensitivity.
Add to favourites

The photons produced by the scintillator crystal need to be amplified and converted into an electrical signal. The diagram below shows the internal structure of a single photomultiplier tube from a gamma camera:

A diagram illustrating a photoelectron system with labeled components: Photoelectron, Visible light photon, Photocathode, Glass tube, Vacuum, Dynodes, and voltage levels ranging from +100 V to +1000 V, with a reference to 0 V.
  1. A single visible light photon from the scintillator hits the photocathode to produce a single photoelectron.
  2. The photoelectron is then accelerated towards the first dynode, an electrode with a potential difference of .
  3. When the photoelectron collides with the dynode, it produces additional electrons known as secondary electrons.
  4. These are then accelerated to the next dynode, which is held at a higher potential difference. Each electron produces, on average, four secondary electrons at each dynode – creating an exponential growth of electrons.
  5. The electrons are all collected at an anode at the end, producing a current that passes through a resistor and generates a small voltage pulse.
Add to favourites

Some of the advantages and disadvantages of using a gamma camera for medical imaging are listed below:

Advantages:

  • Non-invasive imaging method.
  • Provides functional information about the organs, whereas X-rays and CT scans mainly show structure.
  • Medical tracers can be tailored to target specific tissues or organs, allowing for highly focused imaging of the affected areas.

Disadvantages:

  • Image resolution may be affected by scattered photons from the body or low scintillator efficiency.
  • The patient is exposed to a small dose of radiation.
  • Gamma cameras are expensive to purchase and maintain, and the need for radiopharmaceuticals makes the procedure costly and less accessible.
Add to favourites

Fluorine-18 is a radioactive isotope used as a medical tracer in Positron Emission Tomography (PET) scans.

Fluorine-18 beta-decays and emits a positron , as shown in the nuclear notation below:

Add to favourites

Fluorine-18 decays with a half-life of approximately minutes. So, it must be made in the hospital where the patient is located or at a nearby specialised site with a particle accelerator.

One production method for fluorine-18 is to accelerate protons to high speeds and collide them with oxygen-18 nuclei. This produces fluorine-18 nuclei and neutrons. The process is shown in the nuclear equation below:

Oxygen-18 is non-radioactive and easy to obtain since approximately of natural oxygen consists of this isotope.

Add to favourites

All particles have a corresponding antiparticle with the same mass but opposite charge and opposite values of key quantum numbers (for example, lepton number or baryon number), while other properties such as spin are the same.

When a particle and its corresponding antiparticle collide, they annihilate each other. The mass of both the particle and antiparticle is converted into energy, typically in the form of photons.

The annihilation of a positron and an electron creates two gamma photons. The gamma photons travel in opposite directions in order to conserve momentum, as shown below:

An illustration showing an electron (e-) moving left, a positron (e+) moving down, and two gamma rays (γ) emitted to the right and upward from a central point.
  • If the particle and antiparticle are at rest in the centre-of-mass frame, the photons have equal energy and travel in opposite directions to conserve momentum.
  • If the pair has net momentum in the centre-of-mass frame, the photons are still emitted so that total momentum is conserved, but not necessarily in the opposite direction.
Add to favourites

The image below shows a basic schematic of a PET scanner:

A diagram illustrating a patient lying on a table, surrounded by a ring of gamma detectors. Red arrows indicate the direction of gamma rays emitted from the patient.
  1. The patient lies on a table surrounded by a ring of many gamma detectors. Each gamma detector comprises a sodium iodide scintillator crystal and a photomultiplier tube.
  2. Every gamma photon emanating from the patient’s body that collides with the scintillator crystal produces a voltage pulse.
  3. The ring of gamma detectors can determine the location of the annihilation from the difference in time at which the two gamma rays (gamma photons) strike the diametrically opposite detectors.
  4. The computer receives all the voltage pulses produced at the detectors and analyses them to generate a 3-D image on a display.
  5. The different concentrations of the medical tracer are highlighted in different colours and brightness on the display, allowing the analyser to determine where uptake was greatest or lowest.
Add to favourites

PET scanners may also use carbon monoxide as a medical tracer, with the carbon atom labelled with carbon-11 rather than stable carbon-12. Carbon-11 undergoes beta decay, emitting a positron, and has a half-life of approximately 20 minutes.

Carbon monoxide binds to haemoglobin molecules found in red blood cells. Thus, it can be transported around the body and its concentrations monitored using a PET scanner. This provides information about blood flow in the body, such as a bleed in the brain or obstruction in the heart.

Add to favourites

The majority of PET scanners use the medical tracer fluorodeoxyglucose (FDG). This compound is similar to naturally occurring glucose but has been combined with radioactive fluorine-18. The body treats FDG the same as regular glucose; therefore, when it is injected into a patient, it accumulates most in tissues that have a high respiration rate.

Gamma photons are produced when the positrons emitted from the decay of fluorine-18 annihilate with nearby electrons within the patient’s body. The gamma detectors detect these gamma photons produced via annihilation.

Explain that the gamma photons detected in PET scans come from positron-electron annihilation, NOT directly from the decay of fluorine-18. Recall that each annihilation produces two gamma photons travelling in opposite directions.
Do
State that fluorine-18 directly emits gamma rays, it undergoes beta-plus decay, producing positrons, which then annihilate with electrons to create gamma photons. Confuse positron emission with gamma photon emission. The gamma photons come from annihilation, not the decay itself.
Don't
Add to favourites

Some of the advantages and disadvantages of using PET scans for medical imaging are listed below:

Advantages:

  • PET scanners are noninvasive and can be used to diagnose different types of cancer, monitor brain function, and assist with complex surgeries.
  • PET scans are able to monitor brain function, so they can help identify early onset of brain disorders such as Alzheimer’s disease.
  • Can assess the effectiveness of new medicines and drugs on tissues and organs. They allow researchers and doctors to see how a drug is absorbed, distributed, and metabolised in real time, helping to determine its efficacy and potential side effects before widespread clinical use.

Disadvantages:

  • PET scans are very expensive due to the facilities required to produce the medical tracers necessary for the scans.
  • Typically found only at larger hospitals.
  • Only those patients with complex medical problems are given PET scans.
Add to favourites

Sound waves are a type of mechanical wave. When a sound wave propagates through a medium, the particles oscillate in a direction parallel to the wave’s propagation and energy transfer. Hence, sound waves are categorised as longitudinal waves.

An illustration showing the propagation of wave from a speaker to an ear. The image includes labeled sections for Compression, Rarefaction, and Oscillation of particle, with arrows indicating the direction of wave propagation.

A longitudinal wave is a series of compressions, where the particles are close together, and rarefactions, where the particles are spread out.

Add to favourites

Sound waves can have frequencies ranging from a few hertz (Hz) to multiple gigahertz (GHz).

Humans can only hear sound between and this range is called audible sound.

Sound waves that lie below the audible range (i.e. below ) are called infrasound, and the ones that lie above the audible range (i.e. above ) are called ultrasound.

A diagram illustrating sound waves categorized by frequency. The left section labeled 'Infrasound' shows waves at 20 Hz, the middle section labeled 'Audible sound' shows waves at 20 kHz, and the right section labeled 'Ultrasound' shows waves. The horizontal axis is labeled 'Frequency'.
Add to favourites

The generation and detection of ultrasound waves, mainly used by medical devices, depend on the piezoelectric effect.

The piezoelectric effect is the ability of certain piezoelectric crystals, such as quartz and certain ceramics, to generate an electric potential difference in response to mechanical stress (compression or extension).

The effect is reversible. So, a piezoelectric crystal will deform when a potential difference is applied.

Stress is applied ↔ Potential difference is observed. Compression. Piezoelectrical crystal. Compression. Potential difference is applied ↔ Stress is observed. Compression. Piezoelectrical crystal.
  • When stress is applied to a piezoelectric crystal, mechanical energy is converted into electrical energy.
  • When a potential difference is applied to a piezoelectric crystal, electrical energy is converted into mechanical energy.
Add to favourites

The diagram below shows the molecules of a piezoelectric crystal.

An illustration showing a piezoelectric crystal on the left, with arrows indicating compressing force in the middle, and stretching force on the right.
  • Under no stress, the molecules are randomly oriented and no charge is observed at either end of the crystal.
  • When the crystal is compressed, the molecules orient themselves so that the crystal becomes electrically polarised.
  • However, when the crystal is stretched, the polarisation is reversed.

For example, in the diagram above, compressing the piezoelectric crystal may induce a positive charge at the upper end of the crystal and a negative charge at the lower end, while stretching the crystal has the opposite result.

Add to favourites

When a power source is used to apply an alternating potential difference to the ends of the piezoelectric crystal, it compresses and expands so that the molecules of the crystal align with the external electric field, as shown below.

An illustration showing two states of a material: on the left, 'Compression' with arrows indicating force direction and an electric field vector E; on the right, 'Expansion' with similar arrows and the electric field vector E.
  1. Applying an alternating voltage at frequency to the piezoelectric crystal would compress and expand, or vibrate, the crystal at that frequency.
  2. As the crystal vibrates, it disturbs the surrounding air particles, thus producing sound waves of frequency

By applying an alternating potential difference with a frequency greater than ultrasound waves are produced.

Add to favourites

A device that converts energy back and forth between mechanical energy and electrical energy by utilising the piezoelectric effect is called a piezoelectric transducer.

A piezoelectric transducer that is used in ultrasound imaging is shown below.

An illustration of a medical ultrasound transducer showing labeled parts including Coaxial cable, Insulating case, Backing layer, Piezoelectric crystals, and Acoustic lens.

A transducer consists of a thin slice of piezoelectric crystal protected by an insulating case. An acoustic lens at the front is used to converge the ultrasound beam.

The crystal is backed by a damping layer, usually epoxy resin, to quickly stop its vibrations after emitting the ultrasound pulse. This reduces interference between the emitted and reflected pulses.

To maximise the efficiency of a transducer, the frequency of the waves must match the resonant frequency of the crystal. The thickness of the crystal should be about half the wavelength of the emitted ultrasound.

  • To generate ultrasound, the transducer converts an alternating potential difference into sound waves.
  • To detect ultrasound, the transducer converts the sound waves into an alternating potential difference.
Add to favourites

The simplest type of ultrasound imaging is called an A-scan, or amplitude scan:

  • They utilise a single transducer to emit a signal and subsequently receive the reflected signal.
  • A-scans are one-dimensional, determining the depth of internal structures.

The depth to the subject of the scan from the skin is calculated by measuring the time delay between the signal generation and reception.

An illustration showing an eye with a probe, labeled 'Voltage' on the left side and 'Time' on the bottom, with a graph depicting voltage changes over time.

In A-scans, ultrasound waves are typically reflected at the boundary between two media.

Examples of A-scan usage include:

  • measuring the length of the eye,
  • identifying the presence of tumours.

A-scans provide depth measurements only and do not produce images.

Add to favourites

The depth or distance to a target in the body can be calculated using an A-scan.

For a time delay between the generation and reception of the ultrasound, and by knowing the speed of ultrasound in the tissue, the depth of the target can be calculated from:

The factor of is due to how the ultrasound pulse must travel from the transducer to the boundary of the target and then back again, so the measured time corresponds to twice the one-way distance.

An illustration showing an eye with a probe, labeled 'Voltage' on the vertical axis and 'Time' on the horizontal axis. The graph depicts voltage changes over time with peaks and troughs, and the symbol 'Δt' indicating a time interval.
Add to favourites

A B-scan, also known as a brightness scan, is an ultrasound imaging technique that generates two-dimensional or three-dimensional images of internal body structures. B-scans use a moving sensor or multiple sensors to capture ultrasound waves.

The transducer is moved across the patient’s skin, or multiple transducers are employed to acquire a series of A-scans, which are then combined to create detailed images.

Add to favourites

A B-scan image shows the reflected ultrasound intensity at each point on the sensor; for example, high-intensity ultrasound appears as white, while no reflected signal appears black.

An ultrasound image showing a fetus in a womb, with a heart shape visible.
  • B-scans are primarily used for diagnostic purposes, such as visualising bones, muscles, and organs or monitoring a foetus’s development.
  • To enhance image clarity, pulsed ultrasound waves are employed to ensure that reflected waves are received without interference.
  • Shorter wavelengths are used to provide greater detail, allowing sound waves to navigate around finer internal structures.
Add to favourites

The table below highlights the key differences between an A-scan and a B-scan.

A comparison table showing A-scan and B-scan ultrasound techniques. A-scan: A one-dimensional ultrasound scan that measures the distance to internal structures. Displays a single line of data, showing amplitude versus time. Used to measure the length of the eye's internal structures. Typically employs a stationary transducer that focuses on one point. B-scan: A two or three-dimensional ultrasound scan that creates images of internal structures. Produces a cross-sectional image. Used for visualising organs, tissues, and monitoring foetal development. Involves moving the transducer to capture data at different positions.
Add to favourites

Acoustic impedance is a measure of how much resistance an acoustic wave encounters as it travels through a medium.

The acoustic impedance of a substance is defined as the product of the density of the substance and the speed of the ultrasound in the substance:

The SI unit of acoustic impedance is The quantity is the speed of sound in the given substance, not the speed of light.

Add to favourites

The speed of sound in a material depends on both the elasticity and the density of the material.

  • Gases have low densities, and ultrasound travels more slowly in gases than in other states of matter. Therefore, their acoustic impedance is low.
  • Conversely, solids are generally denser, and ultrasound travels fastest in solids. As a result, the acoustic impedance of solids is high.
A table displaying various substances along with their phase, density (kg m-3), impedance (kg m2 s-1), and speed of sound (ms-1). The substances listed include Air (Gas, 1.23, 414, 343), Nitrogen (Gas, 1.25, 412, 334), Oxygen (Gas, 1.43, 484, 316), Ethanol (Liquid, 789, 1,200, 1,161), Oil (Liquid, 850, 1,300, 1,420), Water (Liquid, 1000, 1,480, 1,482), Aluminium (Solid, 2,700, 70,000, 5,100), Steel (Solid, 7,850, 43,000, 5,960), and Copper (Solid, 8,960, 42,000, 3,850).
Add to favourites

When a collimated beam of ultrasound is incident on a boundary between two media, it is partially reflected, and the rest is transmitted into the next medium.

The ratio of the reflected beam intensity to the incident beam intensity depends on the acoustic impedances, and of Medium 1 and Medium 2.

A diagram illustrating the behavior of light at a boundary between two media. The top section labeled 'Medium 1' shows an 'Incident beam' directed downward, with a 'Reflected beam' directed upward. The bottom section labeled 'Medium 2' shows a 'Transmitted beam' directed downward, with a 'Boundary' separating the two media.
Add to favourites

When the incident ultrasound beam is perpendicular to a boundary between two media, the relationship between the reflected beam intensity incident beam intensity and acoustic impedances, and is given by:

The ratio is called the intensity reflection coefficient.

The SI unit of intensity is

  • The greater the difference between the acoustic impedances of the two media, the greater the reflected intensity, resulting in a smaller portion of the beam being transmitted.
  • If the acoustic impedances are the same, the reflected intensity is zero. This is equivalent to the ultrasound wave travelling through a single material with no boundary.
Add to favourites

The transmitted intensity is equal to the incident intensity minus the reflected intensity

The percentage of the initial intensity that is transmitted, added to the percentage of the initial intensity that is reflected, equals 100%:

Add to favourites

During an ultrasound examination, a transducer is typically placed on the patient’s abdomen or chest. If air pockets are trapped between the transducer and the skin, a significant difference in acoustic impedance is created.

As a result, a large portion of the incident ultrasound beam is reflected back, with only a small amount transmitted into the body. This can make studying the internal structure of the human body difficult.

To counteract this effect, a special gel, known as a coupling gel, is placed between the skin and the transducer. An example of this is shown in the diagram below, during the study of a lesion.

An illustration showing the layers of skin, fat, muscle, and a lesion with a tumour. The transducer is positioned on the skin with coupling gel applied.
Add to favourites

The coupling gel used in ultrasound examinations has an acoustic impedance very close to the value of the skin.

  • It eliminates air gaps and enhances the transmission of ultrasound into the body.
  • When a coupling gel is used, we say we are applying impedance matching.
  • When a coupling gel is used, reflection at the boundary is minimal.

From the equation below, the reflected intensity is equal to zero when Therefore, the closer the values of the acoustic impedances of the two media are, the smaller the reflected intensity will be.

Add to favourites

When ultrasound reflects off a moving object, its frequency changes. This is known as the Doppler effect.

Doppler ultrasound is a non-invasive medical technique that utilises ultrasound reflections from iron-rich blood cells. This helps doctors evaluate blood flow through major arteries and veins, including those in the arms, legs, neck, and heart.

An illustration showing a transducer with a transmitting element and a receiving element positioned above tissue and gel. Below, there is a depiction of blood flow with red blood cells and an angle θ indicated.

Applications of Doppler ultrasound:

  • Reveal blood clots.
  • Identify the narrowing of vessel walls resulting from the accumulation of fatty deposits.
  • Evaluate blood flow to transplanted kidneys or livers.
Add to favourites

To perform a Doppler ultrasound:

  1. A transducer is gently placed on the skin above the blood vessel.
  2. The transducer transmits ultrasound pulses and captures the reflected signals.
  3. Reflections from moving blood cells show a change in frequency – the frequency increases when blood flows toward the transducer and decreases when it flows away.
  4. The frequency shift is directly proportional to the speed of the blood in the vessel.
  5. The transducer is connected to a computer that creates a colour-coded image, illustrating the direction and speed of blood flow on a display.
Add to favourites

In Doppler ultrasound, the change in frequency as a fraction of the initial frequency is equal to:

Where:

  • is the speed of blood,
  • is the speed of the ultrasound, and
  • is the angle between the ultrasound receiver and the line along which the blood is travelling.
An illustration showing a transducer with a transmitting element and a receiving element positioned above a blood vessel. The image includes labels for tissue, gel, and blood flow, with an angle θ indicated in relation to the blood flow.
Add to favourites